Details
Features
• Structural reinforcements
• All worn and broken parts replaced
• Current upgrades and FMI’s
• Column +/- 270 degree rotation
• Horizontal arm rebuilt
• All bearings replaced as necessary
• Rebuilt Base assembly
• Latch upgrade
• Upgraded Charger Board
• AMX 4 Plus HV cables
• AMX 4 Plus main cables
• Noise suppressors
• Base anti-static grounding cable
• New hand switch with new cable
• Chassis reinforcement
• New collimator cable
• New rotor cable
• New Super Flex HV Cables
• Collimator rebuilt
• Cross-hair window replaced
• New arm lock
• New batteries charge cycled
• Main frame chassis reinforcement
installed
• All bearings replaced as necessary
• New bumper pads
• New hospital grade cord cap with cord
• New 28 Amp/hour batteries
• Factory set light to x-ray calibration
• Variable speed drive calibrated
• Complete system calibration
• Complete cosmetic repairs and
refinishing
• Complete service documentation
Technical Specifications
• High frequency generator.
• GE AMX-4 Collimator
• 100 mA constant.
• 24 kVp stations.
• kVp 50-125.
• 30 mAs stations.
• mAs 0.4-320.
• LED display.
• Rotating Anode X-Ray tube
• MX-75 275kHU Focal spot .75mm
• Variable speed motor drive
• Electric solenoid locks